Single-use Temp Data Logger

Temperature Humidity Data Logger-Single use

Temperature data logger is a product developed by Apresys specifically for the cold chain industry. It is widely used in food, medicine, logistics, blood, chemical, electronics and other industries. After the data recording is completed, the device is plugged into the computer‘s USB interface to read, save, print data.
Online inquiry
  • Content details

Specifications

Measurement range: -30 ℃ ~ 70 ℃ (-22 ℉ ~ 158 ℉), 0% to 100% RH

Accuracy: ± 0.5 ℃, ±3%RH

Resolution: 0.1℃, 0.1%RH

Data capacity: 24000 readings (12000 temperature data & 12000 RH data)

Interval time: 20 min

Start-up delay: 30 min

Start: Press START button

Stop: Connect to PC or when memory is full (STOP button doesn`t work)

Report Type: PDF format

Alarm: No high and low limits (Optional)

Operating System: WIN XP / VISTA / 7 / 8 / 10 /MAC

Power Supply: ER14250 3.6V lithium battery

Dimensions: 100 mm x 48 mm x 19 mm

Weight: 50 g

 

Operating Instructions

1. To start the data logger, press and hold the START button for 5 seconds, the LED light flashes slowly 2 times then rapidly 10 times. The REC will be displayed in the upper left corner of the screen, indicating that the data logger starts working and recording data.

2. If the memory is full, the data logger will stop working automatically.

3. During recording, if the data logger is connected to a computer, it will automatically stop recording and generating a data report.

4. Do not disconnect the computer while the report is being generated. Otherwise, the data will be lost.

Note:

If the data report is lost, press and hold the START button for 10 sec, it will re-generate a new report.

 